Human C1QTNF12 (C1q and Tumor Necrosis Factor Related Protein 12)  ELISA Kit

Human C1QTNF12 (C1q and Tumor Necrosis Factor Related Protein 12)  ELISA Kit

Size: 96T Price: $420
Reactive Species: Human
Assay Range: 78.13-5000pg/mL
Sensitivity: 46.88pg/mL
SKU: XEH3327 Category: Human Tags: , ,

Datasheet

INFORMATION

Full NameC1q and Tumor Necrosis Factor Related Protein 12
SynonymC1QTNF12, C1QDC2, CTRP12, FAM132A, family with sequence similarity 132 member A, C1q and TNF related 12
Assay TypeSandwich ELISA
ReactivityHuman
Range78.13-5000pg/mL
Sensitivity46.88pg/mL
Sample TypeSerum, plasma and other biological fluids
Sample Volume100μL
Detection WavelengthOD450
Reaction Duration3.25h
Transportation Temperature2-8°C
SpecificitySpecifically binds with C1QTNF12, no obvious cross reaction with other analogues.
Microplate96-wells plate breakable into 12 x 8 wells strip

ELISA KIT COMPONENTS

Upon receipt, unpack promptly and store as recommended in the instructions.

ComponentsSpecificationsStorage and Notes
Micro Plate
96T: 8 wells×12 strips
48T: 8 wells×6 strips
Unopened: -20°C, 12 months
Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C.
Reference Standard
96T: 2 vials
48T: 1 vial
Unopened: -20°C, 12 months
Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution.
Biotinylated Detection Ab Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C, 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
HRP Conjugate Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C(Protect from light), 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
Biotinylated Detection Ab Diluent
14mL×1
2-8℃, 12 months
HRP Conjugate Diluent
14mL×1
2-8℃, 12 months
Reference Standard & Sample Diluent20mL×1
2-8℃, 12 months
Washing Buffer Concentrate (25×)30mL×1
2-8℃, 12 months
Substrate Reagent(TMB)10mL×12-8°C(Protect from light),12 months
Stop Solution7mL×12-8°C/Room temperature

REQUIRED INSTRUMENTS AND REAGENTS

  • Microplate reader (450nm wavelength filter).
  • Incubator capable of maintaining 37℃.
  • 1.5mL EP tube, absorbent paper.
  • Pipette and disposable tip: 0.5-10μL, 2-20μL, 20-200μL, 200-1000μL.
  • Double distilled water or deionized water.

ASSAY PROCEDURE SUMMARY

1
Add 100μL standard or sample to each well, incubate at 37℃ for 90 min. Discard the liquid.
2
Add 100μL Biotinlated Detection Ab working solution to each well. Incubate at 37℃ for 60 min.
Washing: Aspirate and wash the plate for 3 times.
3
Add 100μL HRP conjugate working solution. Incubate at 37℃ for 30 min.
Washing: Aspirate and wash the plate for 5 times.
4
Add 90μL Substrate Reagent. Incubate at 37℃ for 15 min.
5
Add 50μL Stop Solution.
6
Readings at 450nm. Calculation of results.

STANDARD CURVE

The product has been verified by the quality inspection department and meets the performance requirements of the product manual. (Laboratory temperature: 20-25℃, laboratory relative humidity: 20-60%. Before the experiment, all components were equilibrated to room temperature.) The following data is for reference only. A standard curve must be run with each test.
STD.(pg/mL)OD-1OD-2Average
00.0530.0630.058
78.130.1050.1230.114
156.250.1780.1580.168
312.50.2910.2870.289
6250.4680.4980.483
12501.0571.0271.042
25001.6241.6641.644
50002.3912.4352.413
 Standard Curve Image

RECOVERY

5 different levels of target protein spiked in samples were evaluated for recovery experiments to determine the range of recovery and calculate the average recovery rate.

LINEARITY

5 different levels of target protein spiked in samples were evaluated for recovery experiments to determine the range of recovery and calculate the average recovery rate. Subsequently, the 5 samples were diluted to 2-fold, 4-fold, 8-fold, and 16-fold 10 concentrations, and recovery experiments were performed again to evaluate the range and average recovery rate at different dilution factors.

PRECISION(%)

Intra-assay precision (Precision within an assay): Low, mid range and high levels of the samples were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): Low, mid range and high levels of the samples were tested on 3 different plates, 20 replicates in each plate, respectively.

ItemIntra-assay PrecisionInter-assay Precision
Sample123123
n202020202020
Mean (pg/mL)227.28691.322190.03235.72638.272045.75
Standard deviation11.4128.97113.8811.9333.3278.56
CV(%)5.024.195.25.065.223.84

STABILITY

Perform stability testing of the kit at 37°C and 2-8°C and obtain relevant data.

ELISA kit(n=5)37°C for 1 month2-8°C for 6 months
Average(%)8095-100