Human CRP-Ab (C Reactive Protein Antibody) ELISA Kit

Human CRP-Ab (C Reactive Protein Antibody) ELISA Kit

SKU: XEH3163
Size: 96T Price: $320
Reactive Species: Human
Assay Range: 7.81-500ng/mL
Sensitivity: 4.69ng/mL
Category: Human Tags: ,

INFORMATION

Full NameC Reactive Protein Antibody
SynonymCRP-Ab
Assay TypeSandwich
ReactivityHuman
Range7.81-500ng/mL
Sensitivity4.69ng/mL
Sample TypeSerum, plasma and other biological fluids
Sample Volume100μL
Detection WavelengthOD450
Reaction Duration3.25h
Transportation Temperature2-8°C
SpecificitySpecifically binds with CRP-Ab, no obvious cross reaction with other analogues.
Microplate96-wells plate breakable into 12 x 8 wells strip

ELISA KIT COMPONENTS

Upon receipt, unpack promptly and store as recommended in the instructions.

ComponentsSpecificationsStorage and Notes
Micro Plate
96T: 8 wells×12 strips
48T: 8 wells×6 strips
Unopened: -20°C, 12 months
Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C.
Reference Standard
96T: 2 vials
48T: 1 vial
Unopened: -20°C, 12 months
Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution.
Biotinylated Detection Ab Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C, 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
HRP Conjugate Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C(Protect from light), 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
Biotinylated Detection Ab Diluent
14mL×1
2-8℃, 12 months
HRP Conjugate Diluent
14mL×1
2-8℃, 12 months
Reference Standard & Sample Diluent20mL×1
2-8℃, 12 months
Washing Buffer Concentrate (25×)30mL×1
2-8℃, 12 months
Substrate Reagent(TMB)10mL×12-8°C(Protect from light),12 months
Stop Solution7mL×12-8°C/Room temperature

REQUIRED INSTRUMENTS AND REAGENTS

  • Microplate reader (450nm wavelength filter).
  • Incubator capable of maintaining 37℃.
  • 1.5mL EP tube, absorbent paper.
  • Pipette and disposable tip: 0.5-10μL, 2-20μL, 20-200μL, 200-1000μL.
  • Double distilled water or deionized water.

ASSAY PROCEDURE SUMMARY

1
Add 100μL standard or sample to each well, incubate at 37℃ for 90 min. Discard the liquid.
2
Add 100μL Biotinlated Detection Ab working solution to each well. Incubate at 37℃ for 60 min.
Washing: Aspirate and wash the plate for 3 times.
3
Add 100μL HRP conjugate working solution. Incubate at 37℃ for 30 min.
Washing: Aspirate and wash the plate for 5 times.
4
Add 90μL Substrate Reagent. Incubate at 37℃ for 15 min.
5
Add 50μL Stop Solution.
6
Readings at 450nm. Calculation of results.

STANDARD CURVE

The product has been verified by the quality inspection department and meets the performance requirements of the product manual. (Laboratory temperature: 20-25℃, laboratory relative humidity: 20-60%. Before the experiment, all components were equilibrated to room temperature.) The following data is for reference only. A standard curve must be run with each test.
STD.(ng/mL)OD-1OD-2Average
00.080.0980.089
7.820.1320.150.141
15.630.1990.1830.191
31.250.3130.2870.3
62.50.4760.4840.48
1250.9860.9560.971
2501.5961.6461.621
5002.3692.3992.384
 Standard Curve Image

RECOVERY

The recovery of Human CRP-Ab spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.

LINEARITY

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Human CRP-Ab in various matrices were diluted with the Reference Standard & Sample Diluent to produce samples with values within the dynamic range of the assay.

PRECISION(%)

Intra-Assay Precision (Precision within an assay): Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays): Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.

ItemIntra-assay PrecisionInter-assay Precision
Sample123123
n202020202020
Mean (ng/mL)26.7142.13194.526.1838.5176.92
Standard deviation1.352.388.831.741.627.01
CV(%)5.045.654.546.644.213.96

STABILITY

Perform stability testing of the kit at 37°C and 2-8°C and obtain relevant data.

ELISA kit(n=5)37°C for 1 month2-8°C for 6 months
Average(%)8095-100